Detail of Recall

Type of Product:Device
Medsafe Reference:26072
Brand Name:Measurement Modules
Model:Philips M3015A (Model No. 362393) with a specific hardware version manufactured between 2004 and 2011
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:Affected M3015A extensions are incompatible with the IntelliVue X3 Patient Monitor (Model No. 367030) and IntelliVue MMX multi-measurement server. If the extension is connected to an X3 or MMX, which is in turn connected to a host monitor, communication between the X3/MMX and host monitor could be disrupted, resulting in a loss of monitoring and one of the following two technical alarms on the host monitor.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/5/2020