Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26072 |
| Brand Name: | Measurement Modules |
| Model: | Philips M3015A (Model No. 362393) with a specific hardware version manufactured between 2004 and 2011 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medizin Systeme Boblingen GmbH |
| Issue: | Affected M3015A extensions are incompatible with the IntelliVue X3 Patient Monitor (Model No. 367030) and IntelliVue MMX multi-measurement server. If the extension is connected to an X3 or MMX, which is in turn connected to a host monitor, communication between the X3/MMX and host monitor could be disrupted, resulting in a loss of monitoring and one of the following two technical alarms on the host monitor. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/5/2020 |