Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26029 |
| Brand Name: | ABL800 Blood Gas Analyser |
| Model: | ABL800 with FLEXQ sampler tray - All Serial Numbers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | The manufacturer has become aware that there is a very low potential risk of patient mix-up on ABL800 analysers with FLEXQ sampler tray. The mix-up will only occur if the analyser is setup in a specific manner and a specific sequence of events occurs. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/5/2020 |