Detail of Recall

Type of Product:Device
Medsafe Reference:26028
Brand Name:EliA Stool Extraction Kit
Model:
Affected:EliA Stool Extraction Kit - Lot numbers 0236, 0237 and 0238 and EliA Stool Extraction Kit 2- Lot number 0293, 0295, 0299, 0300, 0301, 0302,0303, 0304, 0305, 0306, 0307 and 0309
Software version:
Recalling Organisation:Abacus dx New Zealand
Unit B, 30 Vestey Drive
Mt Wellington
Auckland 1060
Contact Information:0800222170
Manufacturer:Phadia AB
Issue:The manufacturer has received a customer complaint that EliA Stool Extraction Kit 2 had only three (instead of four) notches on the dipstick rod. Investigations showed that there could be other error modes such as missing or broken dipstick rods or missing notches. If a broken dipstick rod is used for stool sampling, it may lead to falsely low or negative results
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/5/2020