Detail of Recall

Type of Product:Device
Medsafe Reference:26024
Brand Name:Drager VentStar Disposable Breathing circuits
Model:
Affected:Part Number: MP02606
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:The manufacturer has become aware in which a VentStar Helix Heated Breathing Circuit was used together with a Fisher & Paykel MR810 humidifier; this product combination is not approved by the manufacturer.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/5/2020