Detail of Recall

Type of Product:Device
Medsafe Reference:26000
Brand Name:IMMAGE Immunochemistry Systems
Model:C3, C4, HPT, IGA, and IGM Reagents; PNs 446450, 446490, 446480, 446460, 447610
Affected:
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has identified that the following chemistries fail to meet IFU interference claims at high levels of Lipid (Triglycerides): C3 at 1000 mg/dL, C4 at 500 mg/dL, HPT at 750 mg/dL, IGA at 1000 mg/dL and IGM at ?250 mg/dL.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/9/2020