Detail of Recall

Type of Product:Device
Medsafe Reference:25997
Brand Name:Biomet Trauma - Instrument Tray I
Model:
Affected:ALL Lots
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Biomet Trauma
Brand Name:CEP Swiss Instruments
Model:
Affected:ALL Lots
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Zimmer GmbH
Issue:The manufacturer is conducting a medical device recall for all lots of the Outer Sheath and the 6.5/8.0MM Screw/Instrument Tray. The affected device and tray that houses the device did not pass steam sterilization process validation testing. The affected device and tray are being obsoleted. Please note that all other instruments in this Tray, excepting the two recalled items have completed the Cleanability and Sterilization Review and these were the only two instruments that failed.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/5/2020