Detail of Recall

Type of Product:Device
Medsafe Reference:25985
Brand Name:ARCHITECT iGentamicin Reagent Kit
Model:
Affected:All Lots
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Diagnostics Division Ireland
Issue:The manufacturer identified that samples tested for ARCHITECT STAT High Sensitive Troponin-I and STAT High Sensitivity Troponin-I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted. Preliminary results from Abbott internal analysis of field data from these two assays show false elevation of patient results with a frequency of 0.002%.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/5/2020