Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25978 |
| Brand Name: | Access Sensitive Estradiol Calibrators |
| Model: | |
| Affected: | Access Sensitive Estradiol Calibrators: 989909,989910, 989911, 989912. Access Free T3 Calibrators:989909, 989910, 989911, 989912 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has discovered that four active Access Sensitive Estradiol calibrator (P/N B84494) lots and four expired Access Free T3 calibrator (P/N A13430) lots share the same lot numbers. The system prevents an operator from configuring a Sensitive Estradiol calibrator lot if an expired Free T3 calibrator with a shared lot number is also configured. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/4/2020 |