Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25976
Brand Name:FluoroQuench™
Model:
Affected:Catalog ID: FQAE100 Lot number: 047A Batch: 0000336132 Expiration Date:10/05/2020 and Catalog ID: FQAE500 Lot number: 047A Batch: 0000336129 Expiration Date: 10/06/2020
Software version:
Recalling Organisation:Bio-Strategy Ltd
22a William Pickering Drive
North Harbour
Albany
Auckland 0751
Contact Information:09 969 9150
Manufacturer:One Lambda Inc
Issue:FluoroQuench™ (Catalog ID FQAE100 and FQAE500) contains particles in the solution. These particles appear as specks in final imaging of the test result. There is low risk to patient or end user as a result of these particles. There is no impact to product as results can still be determined as positive or negative. Test results should be further investigated by the HLA specialist, as part of secondary and confirmatory testing as determined by ASHI and EFI guidelines.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/4/2020
Hide menus