Detail of Recall

Type of Product:Device
Medsafe Reference:25975
Brand Name:Radiometer ABL90Flex; ABL80Flex; ABL9 analysers
Model:ABL80 series analysers - All software versions
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The manufacturer has become aware that the internal barcode reader, included in all versions of the ABL80 analyser, may potentially misinterpret the contents of a locally printed barcode label used for entering patient ID or accession number into the analyser in connection with a sample measurement. The issue identified only relates to barcode types not using a check digit. A check digit enables the barcode reader to validate the data read, and hence, to capture if the barcode has been misinterpreted and consequently to reject such data.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/4/2020