| Type of Product: | Device |
| Medsafe Reference: | 25975 |
| Brand Name: | Radiometer ABL90Flex; ABL80Flex; ABL9 analysers |
| Model: | ABL80 series analysers - All software versions |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | The manufacturer has become aware that the internal barcode reader, included in all versions of the ABL80 analyser, may potentially misinterpret the contents of a locally printed barcode label used for entering patient ID or accession number into the analyser in connection with a sample measurement. The issue identified only relates to barcode types not using a check digit. A check digit enables the barcode reader to validate the data read, and hence, to capture if the barcode has been misinterpreted and consequently to reject such data. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/4/2020 |