Detail of Recall

Type of Product:Device
Medsafe Reference:25949
Brand Name:Angiographic X-Ray
Model:Product code - System description: 722079 - Azurion 7 M20
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message is displayed in the main screen directing the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/4/2020