Type of Product:Device
Medsafe Reference:25945
Brand Name:Patient monitor module, multifunction
Model:
Affected:- ECG malfunction
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:The CARESCAPE ONE monitor may lose ECG parameter and waveform data when used in conjunction with an Automated External Defibrillator (AED) or any AED capable defibrillator when used in AED mode. If this situation occurs, the CARESCAPE ONE will display "ECG Malfunction" and "Check Device" technical alarms. This issue can result in delayed clinical assessment of a life-threatening ECG arrhythmia event.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/4/2020