Type of Product:Device
Medsafe Reference:25938
Brand Name:Access hsTnI Reagent
Model:
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that testing a sample with cardiac troponin I (cTnI) concentrations that are >270,000 pg/mL (ng/L) may cause carryover into the hsTnI reagent pack. The Access hsTnI quantifiable range is up to ~270,000 pg/mL (ng/L).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2020