Detail of Recall

Type of Product:Device
Medsafe Reference:25937
Brand Name:Optilite Analyser IE700
Model:
Affected:Version 7.0, Version 7.0.1
Software version:
Recalling Organisation:The Binding Site Limited
Grant Thornton New Zealand Ltd
L4, 154 Fanshawe Street
Auckland 1010
Contact Information:
Manufacturer:The Binding Site Group Ltd
Issue:A software bug which might affect Optilite analysers running software version 7.0 or 7.0.1 has been identified.
The software bug is not assay specific. If the bug occurs, it may cause result to be calculated against a previously validated historical calibration curve of the same assay from a different lot. Therefore, there is a potential to affect result accuracy.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2020