Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25937 |
| Brand Name: | Optilite Analyser IE700 |
| Model: | |
| Affected: | Version 7.0, Version 7.0.1 |
| Software version: | |
| Recalling Organisation: | The Binding Site Limited Grant Thornton New Zealand Ltd L4, 154 Fanshawe Street Auckland 1010 |
| Contact Information: | |
| Manufacturer: | The Binding Site Group Ltd |
| Issue: | A software bug which might affect Optilite analysers running software version 7.0 or 7.0.1 has been identified. The software bug is not assay specific. If the bug occurs, it may cause result to be calculated against a previously validated historical calibration curve of the same assay from a different lot. Therefore, there is a potential to affect result accuracy. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/4/2020 |