Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25936 |
| Brand Name: | HumaTrol P |
| Model: | |
| Affected: | Lot - 0004 |
| Software version: | |
| Recalling Organisation: | Thermo Fisher Scientific New Zealand Ltd 244 Bush Road Albany AUCKLAND |
| Contact Information: | 0800 636 327 |
| Manufacturer: | Human Gesellschaft GmbH |
| Issue: | The manufacturer has advised that recent lots of AUTOCAL and HUMAN control materials exhibit deviations of approximately 20% for ALKALINE PHOSPHATASE opt. liquicolor (DGKC) when traced back to the molar extinction coefficient. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/4/2020 |