Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25935
Brand Name:Blood Collection Tube, non evacuated, sterile
Model:
Affected:Lot - 191101
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:Greiner Bio-One GmbH (Austria)
Issue:The manufacturer has become aware that MiniCollect blood collection tubes, partly show a wrong blood to additive ratio. According to ISO 6710 +/- 10% tolerance, the acceptable range is between 1:7,36 to 1:11.
If an affected LOT is used for blood sample analysis, blood to additive ratio might be < 1:7 and the test results might show a prolonged aPTT.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2020