Detail of Recall

Type of Product:Device
Medsafe Reference:25912
Brand Name:AQT90 series analysers
Model:This relates to barcode types not using a check digit
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The manufacturer has become aware that the barcode readers in the bottom edge of the display and behind the inlet wheel may misinterpret the contents of a locally printed barcode label used for entering patient ID or accession number into the analyser in connection with a sample measurement. This relates to barcode types not using a check digit.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/4/2020