Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25908 |
| Brand Name: | Medtronic model 2090 programmer |
| Model: | 2090; All Serial Numbers with Software Application SW034 version 8.3 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | CRDM - Carelink Encore Programmer Model 29901 |
| Model: | 29901, 29901A; All Serial Numbers with Software Application SW034 version 8.3 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | The manufacturer has identified that the 2% risk for a partial electric reset during the interrogation process is due to an uncommon scenario when a software update is installed simultaneously with routine critical memory scans. Should a reset occur, the clinician will be prompted to "Clear" the reset condition on the programmer (guidance to clear a partial reset is documented in the Instructions for Use for the above-named devices). When the "Clear" option is selected, the programmer will automatically interrogate the device again, and will successfully write the software enhancement to the device memory. Importantly, 98% of download attempts will successfully complete without an electrical reset. Once the software update has been successfully installed into the device, the potential for a future partial reset due to this interaction no longer exists. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/4/2020 |