Detail of Recall

Type of Product:Device
Medsafe Reference:25902
Brand Name:Accelerator APS Aliquoter Module
Model:
Affected:All serial numbers
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Diagnostics
Issue:The APS Aliquoter's current error recovery procedure dispenses the aliquoter volume back into the Primary Tube in the case of an AQM "clot detection error" (error code E0E0). If the error is not due to a sample clot, but generated from an aliquoter hardware failure, a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/4/2020