Detail of Recall

Type of Product:Device
Medsafe Reference:25892
Brand Name:Afinion 2 analyser and Alere Afinion AS100 Analyser
Model:
Affected:Afinion- catalogue number - 1116679, 1116770 and Alere Afinion™ AS100 Analyzer - catalogue number - 1116049
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Diagnostics Division Ireland
Issue:The manufacturer has received complaints for the Afinion Analyzer systems related to electrostatic discharge and sparking of the power supply provided with the Analyzers. Customers have experienced power supplies becoming non-functional after these events.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/4/2020