| Type of Product: | Device |
| Medsafe Reference: | 25892 |
| Brand Name: | Afinion 2 analyser and Alere Afinion AS100 Analyser |
| Model: | |
| Affected: | Afinion- catalogue number - 1116679, 1116770 and Alere Afinion™ AS100 Analyzer - catalogue number - 1116049 |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Diagnostics Division Ireland |
| Issue: | The manufacturer has received complaints for the Afinion Analyzer systems related to electrostatic discharge and sparking of the power supply provided with the Analyzers. Customers have experienced power supplies becoming non-functional after these events. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/4/2020 |