Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25878
Brand Name:Caesarea Syringe Pump
Model:T34 and T34L pumps
Affected:
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Caesarea Medical Electronics Ltd (Israel)
Issue:The manufacturer BD/CME has identified that the syringe pump motor block mechanism may be affected overtime by "wear / tear". This scenario causes under infusion situations without the pump alarm being generated.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/3/2020
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /