Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25873
Brand Name:Laser Fiber
Model:MoXyTM Laser Fiber
Affected:
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Boston Scientific Corporation
Issue:The manufacturer is initiating a Product Correction for the GreenLight MoXyTM Laser Fibers to reinforce existing instructions within the Directions for Use (DFU) and provide further guidance which will be added to the DFU that may reduce the likelihood of metal cap and fiber tip break temperature related complaints. Internal bench testing has determined that an increased irrigation flow may increase the liquid cooling effect and may reduce temperature related complaints.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/3/2020
Hide menus