Detail of Recall

Type of Product:Device
Medsafe Reference:25868
Brand Name:Ingenia Series
Model:Philips Ingenia CX 1.5T, Ingenia CX 3.0T Ingenia 1.5T, Ingenia 3.0T, Ingenia Ambition, Ingenia Elition, Achieva dStream 1.5T, Achieva dStream 3.0T MR systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has become aware of an anomaly with the Philips Ingenia, Ingenia Ambition, Ingenia Elition, Achieva dStream MR systems. The issue is caused by a part of the cast iron scissor (within the patient support) that touches a cable clamp and forces it downwards. When this cable clamp moves downwards the bar containing the end switches also moves down, resulting that the end switches are in the wrong position. Consequently, the upper position of the patient support cannot be reached anymore. This can be prevented by moving the cable clamp (within the patient support), that holds the bar with the end switches, 2 cm downwards. This must be done by a qualified Philips engineer.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/3/2020