| Type of Product: | Device |
| Medsafe Reference: | 25863 |
| Brand Name: | V60 Ventilator |
| Model: | |
| Affected: | Multiple part numbers |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Respironics California LLC |
| Issue: | A solder connection on the first generation Power Management printed circuit board assembly ((PCBA) P/N 1055906) of affected Philips V60 ventilators is subject to solder connection failure. This solder joint connects a component (designated as R31) to the PCBA. In the most common failure mode of the solder joint, the failure will cause the blower to lose power, spool down, and trigger a visual and audible High Priority Check Vent alarm to alert clinicians to switch the patient to alternative ventilation. This failure mode is referred to as an open failure. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/3/2020 |