Detail of Recall

Type of Product:Device
Medsafe Reference:25863
Brand Name:V60 Ventilator
Model:
Affected:Multiple part numbers
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics California LLC
Issue:A solder connection on the first generation Power Management printed circuit board assembly ((PCBA) P/N 1055906) of affected Philips V60 ventilators is subject to solder connection failure. This solder joint connects a component (designated as R31) to the PCBA. In the most common failure mode of the solder joint, the failure will cause the blower to lose power, spool down, and trigger a visual and audible High Priority Check Vent alarm to alert clinicians to switch the patient to alternative ventilation. This failure mode is referred to as an open failure.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2020