Detail of Recall

Type of Product:Device
Medsafe Reference:25855
Brand Name:LeMaitre_10714_IIa_ Embolectomy Catheter
Model:
Affected:Lots - OTW3190, OTW3821, OTW3958, OTW4073, OTW4197
Software version:
Recalling Organisation:Clinical & Regulatory Services Limited
30 Fulford Road
Havelock North
Hastings 4130
Contact Information:(06) 844 4490
Manufacturer:LeMaitre Vascular Inc
Issue:The manufacturer has detected an increased trend in reports of catheters failing to deflate during use. The issue occurs if the thread windings around the balloon slip under tension and allow the inflated balloon's proximal attachment to slide over the inflation hole. Under continued tension, the tip of the catheter can begin to stretch, which can lead to tip separation.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/3/2020