Detail of Recall

Type of Product:Device
Medsafe Reference:25851
Brand Name:Haemodialysis system - Lundia AB
Model:
Affected:upgrade to version 3
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Gambro Lundia AB
Issue:On September 3, 2018, the manufacturer issued the enclosed Urgent Recall for Product Correction communication (Medsafe Ref #23537) regarding the potential for AK 98 Hemodialysis Devices to generate excessive ultrafiltration (UF) in certain situations where treatment-related alarms occur, or where there is an ultrafilter leak. The manufacturer released software version V3.0 for the AK 98 device that mitigates cases of excessive fluid loss in patients. The manufacturer is in the process of upgrading all AK 98 devices to software version V3.0.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/3/2020