Detail of Recall

Type of Product:Device
Medsafe Reference:25804
Brand Name:ABL800 series analysers
Model:ABL800 series analysers with software versions below V6.19.
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The manufacturer has received a few reports of occurrences where the ABL800 barcode reader has misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyser in connection with a sample measurement. The occurrences relate to barcode types not using a check digit.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/3/2020