Type of Product:Device
Medsafe Reference:25788
Brand Name:Expression Series Patient Monitors
Model:
Affected:Software Revisions: 01.03.00, 01.04.00, 01.05.00 and 01.07.00
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Invivo, a division of Philips Medical Systems
Issue:For MR400 devices with software revisions 01.03.00, 01.04.00, 01.05.00 and 01.07.00, the menu selections for users to access the O2 sensor calibration were changed. These changes were not incorporated into the device's Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with these software revisions.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2020