Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25721
Brand Name:Dimension® Chemistry Products
Model:
Affected:Total Bilirubin - test code - TBI All lot numbers
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has become aware that the United Kingdom Medicines and Healthcare products Regulatory Agency published an alert to healthcare professionals informing them that laboratory tests for bilirubin should be monitored for patients who take the drug eltrombopag due to the potential for discordant results
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/2/2020