Detail of Recall

Type of Product:Device
Medsafe Reference:25716
Brand Name:Emicizumab medicine and factor VIII assays
Model:
Affected:All lots
Software version:
Recalling Organisation:Diagnostica Stago Pty Ltd
Level 3
64 Cashel Street
Christchurch
Contact Information:+64 3 377 4115
Manufacturer:Diagnostica Stago SAS
Issue:The manufacturer has identified when using the STA - Deficient VIII or STA - ImmunoDef VIII reagent there is a potential risk of contamination by Emicizumab of factor VIII assays on the following Stago instruments: STA® Compact, STA® Compact Max, STA-R® Evolution or STA-R® Max.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/2/2020