Detail of Recall

Type of Product:Device
Medsafe Reference:25714
Brand Name:ECG Cables & Leadsets
Model:Philips Efficia Combined Cable/3 or 5 Leadsets (Efficia Combined Cable/Leadsets)
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:During use of Efficia Combined Cable/Leadsets, if the patient required electrical energy for defibrillation or cardioversion, it is possible for some of the energy to be diverted away from the patient's thoracic cavity through the ECG cable. This could result in reduction of defibrillation energy delivered to the patient, or failure to deliver defibrillation energy to the patient.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/2/2020