Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25648
Brand Name:Radiometer ABL90Flex; ABL80Flex; ABL9 analysers
Model:
Affected:Serial numbers - ABL90FLEX , 393-090, ABL90FLEX PLUS, 393-092
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The manufacturer has received a few reports of occurrences where the ABL90 barcode reader has misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. The occurrences relate to barcode types not using a check digit.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/2/2020
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /