Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25642
Brand Name:LTF-160 Pleuravideoscope
Model:Biopsy Forceps - Spray catheter - Hot Biopsy Forceps - FD-6, Injector- Electro Surgical Knife- Electro Surgical Snare - Coagulation electrode
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacture, is requesting that you cease the combined use of the Pleuravideoscope LTF-160 with the endotherapy accessories set, except for Biopsy forceps FB-55CR-1.It has been made aware that whilst there has been an assessment of the physical compatibility between the accessories and the LTF-160, there is no data demonstrating that the subject endotherapy accessories are compatible for use with the LTF-160
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/2/2020
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /