Detail of Recall

Type of Product:Device
Medsafe Reference:25629
Brand Name:Elecsys® CA 19-9
Model:
Affected:Non-reproducible elevated results with reagent lots 416245, 464449,483123, 504743, 525510, 568976,598135, 598140 and 599263
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has received complaints concerning non-reproducible elevated results (high flyers)for lots 416245, 464449, 483123, 504743, 525510 and 568976, Elecsys® CA 19-9 on cobas e801 and cobas e402. The issue is reagent lot specific and not related to the cobas e801 or 402 instrument.The issue can lead to non-reproducible elevated CA 19-9 results and therefore affect clinical interpretation (either result of multiple results is discrepant-high compared to other results of the same sample).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/11/2022