Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25612 |
| Brand Name: | Cardiac Science Automated External Defibrillators |
| Model: | All G3 Elite devices |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Paragon Care Group New Zealand Limited 69 Elizabeth Knox Place Glen Innes AUCKLAND |
| Contact Information: | 0508 654 258 |
| Manufacturer: | Cardiac Science Inc |
| Issue: | The manufacturer has received reports of G3 Elite Automated External Defibrillator devices where the Rescue Ready status indicator displays red and the Service LED is illuminated. The manufacturer's investigation traced the problem to a software anomaly associated with the Daylight Savings Time (DST). |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/1/2020 |