Detail of Recall

Type of Product:Device
Medsafe Reference:25611
Brand Name:Heartware - Ventricular Assist System (HVAD)
Model:HeartWareTM Battery Charger packaged with AC Adapter - All serial numbers, HeartWareTM Battery Charger AC Adapter - All serial nimbers
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:HeartWare Inc
Issue:There is the potential for an HVAD™ System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port. When this occurs, communication circuits between the battery and the Controller may be damaged. If circuit damage occurs, it will be necessary to exchange the Controller to restore communication between the batteries and Controller.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:29/1/2020