Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25576 |
| Brand Name: | Lina Medical - Open-surgery electrosurgical handpiece/electrode, monopolar, single-use |
| Model: | |
| Affected: | Lot numbers - 1923011, 1923050, 1924004, 1924005, 1924015, 1929038, 1929039, 1929055 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Lina Medical ApS |
| Issue: | Hairline fractures in a component could allow electrical current to arc out of the Telescopic Smoke Evacuation Pencil. There is a potential for unintended energy delivery resulting in a burn to the patient and/or user. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/1/2020 |