Detail of Recall

Type of Product:Device
Medsafe Reference:25576
Brand Name:Lina Medical - Open-surgery electrosurgical handpiece/electrode, monopolar, single-use
Model:
Affected:Lot numbers - 1923011, 1923050, 1924004, 1924005, 1924015, 1929038, 1929039, 1929055
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Lina Medical ApS
Issue:Hairline fractures in a component could allow electrical current to arc out of the Telescopic Smoke Evacuation Pencil. There is a potential for unintended energy delivery resulting in a burn to the patient and/or user.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/1/2020