Type of Product:Device
Medsafe Reference:25572
Brand Name:OmniDiagnost system
Model: Velara X-ray generator of the Philips Allura Xper, the Integris, the MultiDiagnost Eleva and OmniDiagnost systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Angiographic X-Ray
Model: Velara X-ray generator of the Philips Allura Xper, the Integris, the MultiDiagnost Eleva and OmniDiagnost systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified that a capacitor inside the convertor of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible anymore. The failed capacitor may produce smoke and a burning odor in the room where the generator is located. If the system is a biplane model, X-ray acquisition is still possible on the other channel. If the system is a monoplane, it cannot be used anymore until the capacitor is replaced.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/1/2020