Detail of Recall

Type of Product:Device
Medsafe Reference:25553
Brand Name:119983 - 40887 - Radiation therapy treatment planning system, application program software - IIb - Eclipse - Smart Segmentation - Vitesse - VariSeed
Model:DRR Images Fail to Recalculate or Update as Intended
Affected:
Software version:Eclipse™ Treatment Planning System version 15.5 and 15.6
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Issue:The manufaturer has discovered an anomaly that may occur with Eclipse versions 15.5 or 15.6. DRR images may not be recalculated and updated following a change to field geometry when more than one user is simultaneously making changes to the plan. This is called a concurrent editing session.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/1/2020