Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25553 |
| Brand Name: | 119983 - 40887 - Radiation therapy treatment planning system, application program software - IIb - Eclipse - Smart Segmentation - Vitesse - VariSeed |
| Model: | DRR Images Fail to Recalculate or Update as Intended |
| Affected: | |
| Software version: | Eclipse™ Treatment Planning System version 15.5 and 15.6 |
| Recalling Organisation: | Varian Medical Systems Australasia Pty Ltd (NZ) Millenium Centre, Level 3, Building C, 600 Great South Road Ellerslie Auckland 1051 |
| Contact Information: | 0800 486 732 |
| Manufacturer: | Varian Medical Systems Inc |
| Issue: | The manufaturer has discovered an anomaly that may occur with Eclipse versions 15.5 or 15.6. DRR images may not be recalculated and updated following a change to field geometry when more than one user is simultaneously making changes to the plan. This is called a concurrent editing session. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/1/2020 |