Detail of Recall

Type of Product:Device
Medsafe Reference:25552
Brand Name:Ysio Max
Model:Ysio Max – 10762470
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Brand Name:Luminos & Multitom Rax
Model:Luminos Agile Max - 10762472, Luminos dRF Max – 10762471
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Brand Name:UROSKOP Omnia Max
Model: UROSKOP Omnia Max – 10762473
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:Issue 1.The manufacturer has received a report of a system collision due to inappropriate use of the bypass key which is overriding the built-in collision control. The bypass key is intended to enable moving an otherwise blocked system only to avert an immediate hazard.
Issue 2. The manufacturer has received a report of an image which was assigned to two different patients.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/1/2020