Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25539 |
| Brand Name: | Healing Abutments/cylinders/caps |
| Model: | Certain® BellaTek® Encode® Healing Abutments Lot numbers - 1228905, 1229062 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | ZimVie Australia Pty Ltd (NZ) 16 Huapai Street Onehunga Auckland 1061 |
| Contact Information: | 0508 122 221 |
| Manufacturer: | Implant Innovations Inc |
| Issue: | On January 13, 2020, you received a communication from Zimmer Biomet Dental regarding a product recall with specific lots of BellaTek® Encode® Healing Abutments for the Certain® Internal Connection. The identified BellaTek Encode Healing Abutments were incorrectly manufactured wherein the orientation flat on the Encode Healing Abutment's occlusal surface is misaligned by 30 degrees. As part of further investigation an additional two lots with the same manufacturing condition were recently identified. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 15/5/2020 |