Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25538 |
| Brand Name: | Physio-Control Inc - IIb - 17882 - Adaptor |
| Model: | |
| Affected: | Part numbers beginning with V15-2. |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Physio-Control Corporation |
| Issue: | In January 2020, the manufacturer initiated a lot-specific Recall for Product Correction for LIFEPAK 15 Monitor/Defibrillator devices that may not deliver a shock after the Shock button on the keypad is pressed. The manufacturer has since identified that additional keypads, which are sold as individual units that could later be installed into LIFEPAK 15 devices, are also included in scope of this correction. As a result, the manufacturer has identified additional affected devices that had these individual keypads installed. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/9/2021 |