Detail of Recall

Type of Product:Device
Medsafe Reference:25526
Brand Name: Immuno Concepts Image Navigator software
Model:Immuno Concepts Image Navigator software Version 3 and 4 (ICINUP).
Affected:
Software version:
Recalling Organisation:Abacus dx New Zealand
Unit B, 30 Vestey Drive
Mt Wellington
Auckland 1060
Contact Information:0800222170
Manufacturer:Immunco Concepts
Issue:The manufacturer has identified an issue with Immuno Concepts Image Navigator software, version 3 and 4. Image Navigator assay runs performed in 2020 cannot be viewed due to the '2020' date. The software runs the Image Navigator instrument and records the data generated, but the data cannot be reviewed unless the data file date is first changed to '2019'. Therefore no data has been lost and no workstations or scans are affected.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/1/2020