Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25524
Brand Name:Gore_Excluder AAA Contralateral Leg Endoprosthesis (Codes PLC & PXC)
Model:GORE EXCLUDER AAA
Affected:
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:W L Gore & Associates Inc
Brand Name:Gore_Excluder AAA Trunk Ipsilatral Leg Endoprosthesis (Codes RLT & RMT)
Model:GORE EXCLUDER Iliac
Affected:
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:W L Gore & Associates Inc
Issue:The manufacturer has received reports of leading end catheter component separations of the EXCLUDER Device. Of the events, 30 reported immediate health consequences and 1 long-term health consequence (pelvic ischemia).
This represents a rate of 0.05% reported complaints of leading end catheter separation events over the last 6 years .
Based on these events, Gore will be updating its Instructions for Use.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/1/2020
Hide menus