Detail of Recall

Type of Product:Device
Medsafe Reference:25510
Brand Name:Tempus LS
Model:
Affected:All Tempus LS on the market within listed s/n range are affected; Serial Numbers: 7021.000100 up to 7021.001215
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Schiller AG
Issue:The manufacturer has identified that under certain circumstances a defibrillation will not be possible. It's possible that an impedance change between patient and device causes an impairment of the impedance measurement. If the impedance is out of a valid range, a defibrillation is not possible.
Issue 2: Unexpected safety discharge
The manufacturer received notice that under certain circumstances the device performs an unexpected safety discharge.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/1/2020