Detail of Recall

Type of Product:Device
Medsafe Reference:25507
Brand Name:Light therapy unit, photo
Model:Giraffe Blue Spot PT Lite Phototherapy Systems
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Ohmeda Medical Inc
Issue:The manufacturer has recently become aware of a problem with some Giraffe Blue Spot PT Lite phototherapy systems built between 2014 and Jan 2018, having a light output falling below the recommended minimum output for the Giraffe Blue Spot PT. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2019