Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25505 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation/RayPlan 7, RayStation/RayPlan 8A, RayStation/RayPlan 8B and RayStation/RayPlan 9A: |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has identified an issue with the options 'Map ROI(s)' and 'Map ROI(s) reversed', only available by right-clicking in the ROI list in the Structure Definition module, that may lead to unintended ROI geometries when mapping ROIs using a rigid-only registration. If the rigid registration is modified after the Structure Definition module was first opened, the modifications will not be taken into account when mapping ROIs between the image sets. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/12/2019 |