Type of Product:Device
Medsafe Reference:25505
Brand Name:RaySearch_Radiation therapy treatment planning system
Model:RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation/RayPlan 7, RayStation/RayPlan 8A, RayStation/RayPlan 8B and RayStation/RayPlan 9A:
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:The manufacturer has identified an issue with the options 'Map ROI(s)' and 'Map ROI(s) reversed', only available by right-clicking in the ROI list in the Structure Definition module, that may lead to unintended ROI geometries when mapping
ROIs using a rigid-only registration. If the rigid registration is modified after the Structure Definition module was first opened, the modifications will not be taken into account when mapping ROIs between the image sets.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2019