Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25500
Brand Name:Angiographic X-Ray
Model:Philips VesselNavigator application when used with the Philips Azurion System R1.2.x and R2.0.x
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:Due to a software defect, when a Digital Subtracting Angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction. The unsubtracted images contain more anatomical information than the DSA, which may not be suitable for verification of the registration of a pre-operative CT/MR with X-ray images during the Live Guidance step.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/12/2019