Detail of Recall

Type of Product:Device
Medsafe Reference:25499
Brand Name:Hamburg DXR
Model:All CombiDiagnost R90 GCF systems are affected
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has identified that the table can be tilted by pressing the corresponding + or - tilt motion buttons (Table Tilt Left/Right) or using the table up/down button (Table Up/Down). If using the Table Up/Down button, the table moves down and tilts back to horizontal position at the same time. When using the Table Up/Down button, the system can experience Error 80, which locks the geometry in that specific state. System operation cannot be restored by system reboot, but requires the intervention of a service engineer.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/12/2019