Detail of Recall

Type of Product:Device
Medsafe Reference:25484
Brand Name:FTD CE-IVD Kits
Model:
Affected:All lots
Software version:
Recalling Organisation:Banksia Scientific Pty Ltd (NZ)
12 Brigade Street
Airport Oaks
Auckland
Contact Information:09 921 2626
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Update information:
investigation has confirmed that for identified products, various performance claims listed in the
current Instructions for Use will not be met (i.e. sensitivity, specificity, LoD, etc.). This creates the potential for increased instances of erroneous results (false positives, false negatives, etc.) and confirms a previously unidentified risk to patient health, starting with the date when these kits were first made available
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/2/2020